Model 1040C/CU Panel Meter Calibrator - DISCONTINUED

Model 1040C/CU Panel Meter Calibrator - DISCONTINUED

*Repair and calibration services available through 2022 depending upon parts availability.

*Accessories available until supplies run out.

 

Model 1040C/CU Manual
Download PDF(733 kB) »

Model 1040C/CU Data Sheet
Download PDF(402 kB) »

Model 1040C/CU Data Sheet-Español
Download PDF(146 kB) »

The Arbiter Systems®, Inc. Model 1040C and 1040CU Panel Meter Calibrator (PMC) is discontinued. We will provide repair and calibration services through 2022 depending upon parts availability.

 

 

 

 

The accessories listed below are still available for purchase.

  • Calibrates panel meters, transducers, circuit breakers and overcurrent relays
  • Rugged, portable for field work
  • 1040C includes hand-held controller for remote/distance operations
  • Works with external standard Model 931A
  • Self-test and diagnostics built-in
  • Supports complex measurement sequences with included store-recall memory function

Output Signals

Voltage

Output Range

10 mVdc to 1050 Vdc
1.5 Vrms to 750 Vrms

Accuracy

± (0.2 % setting + 0.05 % FS) (dc)

≤ 150 Vrms ± (0.2 % setting + 0.05 % FS)
> 150 Vrms ± (0.2 % setting + 0.1 % FS)

Note: Internal voltage ranges have full-scale (FS) values of 105 mV, 1.05 V, 10.5 V, 105 V, and 1050 V (dc), and 15.75 Vrms, 157.5 Vrms, and 750 Vrms (ac).

Burden

15 mA; ≥ 25 mA overload (dc)

≤ 150 Vrms 300 mArms
> 150 Vrms 10 VA

Noise (dc)

≤ 0.25 % setting; 10 kHz BW

Distortion

≤ 0.45 %

Auxiliary Voltage

Synchroscope Output is connected to the Voltage Output with a selectable 0 ° or 180 ° phase relationship. The total burden of both outputs must not exceed specification.

Current

Output Range

0.1 mA to 10.5 Adc
0.1 Arms to 7.5 Arms

Accuracy

± (0.2 % setting + 0.05 % FS) (dc)

≤ 1.05 Arms ± (0.2 % setting + 0.05 % FS)
1.05 ≤ 5.25 Arms ± (0.2 % setting + 0.1 % FS)
> 5.25 Arms specification not verified

Note: Internal current ranges have full-scale (FS) values of 1.05 mA, 10.5 mA, 105 mA, 1.05 A, and 10.5 A (dc), and 1.05 Arms, and 7.5 Arms (ac).

Compliance

Vrms; 6.5 V overload (ac)

≤ 50 mAdc 12 Vdc; 12.5 V overload
> 50 mAdc 3 Vdc; 3.5 V overload

Noise (dc)

≤ 0.25 % setting; 10 kHz BW

Distortion

≤ 0.45 %

Frequency - AC Operation

Ranges

50 Hz to 75 Hz
333.3 Hz to 500 Hz

Accuracy

≤ 0.01 %

General

Stability

≤ (0.03 % setting + 0.015 % FS) (see Notes under the Voltage and Current headings) averaged one minute or longer

Resolution

≤ 0.1 % setting

Settling Time

8 seconds maximum

AC Power, VARs

Range

1.5 VA to 5625 VA

Current

0.1 Arms to 7.5 Arms

Voltage

15 Vrms to 750 Vrms

Phase Angle

+ 180 ° to - 180 °

Phase Accuracy

± 0.33 °

Stability

≤ (0.06 % setting + 0.03 % FS) averaged one minute or longer.

Note: Full-Scale (FS) value determined from full-scale values of voltage and current ranges being used.

Other ac power specifications are derived from the individual specifications for current and voltage.

Phase , Power Factor

Phase Angle

+ 180 ° to - 180 °

Phase Accuracy

± 0.33 °

Phase Stability

≤ 0.2 ° rms, averaged one minute or longer

Resolution

≤ 0.1 °

Interface

Operator

Display

20-character alphanumeric VFD

Display Mode

Normal, deviation(%), reset

Keypad

Front-panel numeric keyboard
Up/down increment keys
Digital knob

Instrument Mode

DC, 60 Hz, and 400 Hz

Function

Voltage, current, synchroscope, power, reactive power, frequency, power factor, & phase

Output

Standby (outputs disabled)
Operate (outputs active)

Connectors

Detachable, 3 m cables (1040C)
5-way binding posts (1040CU)

Synchroscope

0 ° or 180 ° (selectable)

Hand-Held Unit

(1040C only) Store, recall, up, down, operate, standby, normal and % deviation keys.

System Interface

Computer

IEEE-488

Power Requirements

External

Range

115 Vrms ± 10 %, 47 Hz to 500 Hz
0.5 A (standby); 3 A (operate) maximum

General

Physical

Instrument/Size

38 cm x 38 cm x 28 cm (15 in x 15 in x 11 in)
533 mm x 533 mm x 533 mm (21 in x 21 in x 21 in), shipping

Weight

17 kg (38 lbs), net
20.4 kg (45 lbs) self, shipping
25 kg (55 lbs) with manuals and cables, shipping

Environmental

Temperature

Operating: 0 °C to 55 °C
Nonoperating: - 40 °C to + 75 °C
Cooling:
Forced air

There are no additional Options available for this product.
Available
AS0000300 1 Meter Adapter 1/4-20 (4 ea)
AS0000400 1 Meter Adapter 10/32 (4 ea)
AS0011303 2 1040C Hand-held control unit with 3 m cable
AS0012900 1 Service Kit
AS0020100 3 1040C/1040CU Maintenance Manual
AS0020200 4 1040C/1040CU Operation Manual
AS0058500 5 Hand-Held Control Unit for 1040CU
CA0005606 6 6 m 1040C Voltage Cable
CA0005703 2 3 m 1040C Current Output Cable
CA0005706 6 6 m 1040C Current Output Cable
CA0011103 6 3 m 1040C Hand-Held Control Unit Extension Cable
P09R 4 Power Cord, North America, Right Angle

1 Available for both Model 1040C and 1040CU
2 Included with Model 1040C
3 Included with Model 1040C, Available for Model 1040CU
4 Included with both Model 1040C and 1040CU
5 Available for Model 1040CU Only
6 Available for Model 1040C only

No software downloads available for this product
 
No additional software information is available for this product
No firmware Downloads are available for this product
 
No additional firmware information is available for this product

Training: Calibrating Meters and Transducers with Model 1040C/CU
Download 1040c_training2_calibrating_meters_transducers.pdf (270 kB) »

Training: Model 1040C Panel Meter Calibrator
Download 1040c_training1.pdf (344 kB) »

Quick Model Search


phone
(US, Canada, Mexico only) 800.321.3831 +1.805.237.3831

email
sales@arbiter.com

Find a Local Representative